Temperature Converter | RSD-TI2-Ex8

Артикул: RSD-TI2-Ex8
Производитель: Pepperl-Fuchs



  • 8-channel
  • Inputs EEx ia IIC
  • Device installation in Zone 1, Zone 2, or Zone 22
  • Module can be exchanged under voltage in Zone 1 (hot swap)
  • Connection of resistance thermometers Pt100 or Ni100 in 2-, 3- or 4-wire connection
  • Connection of thermocouples, type B, E, J, K, L, N, R, S or T
  • Connection of other resistive sensors, mV sources or thermocouples possible. Linearization adjustable via software
  • Lead breakage (LB) monitoring for each field circuit
  • Filter cut-off frequency can be programmed from 0.2 Hz to 10 Hz
  • EMC acc. to NAMUR NE 21

Datasheet as PDF in following languages:


Technical Data: RSD-TI2-Ex8

Supply
Connectionterminals 34, 50 V+; 35, 51 V-
Rated voltage8.88 ... 9.5 V
Power loss1.6 W
Power consumption1.6 W
Internal bus
Connectionbackplane bus
Interfacemanufacturer specific bus
Cycle time1.6 ms
Input
Connectionterminals 0+, 1H, 2L, 3-; 4+, 5H, 6L, 7-; 8+, 9H, 10L, 11-; 12+, 13H, 14L, 15-; 17+, 18H, 19L, 20-; 21+, 22H, 23L, 24-; 25+, 26H, 27L, 28-; 29+, 30H, 31L, 32-
Input signalsuitable for:
Pt100, Pt200, Ni100, Ni200, Ni120 Minco, Cu10 Minco
thermocouples type B, E, J, K, N, R, S, T and mV sources
Lead resistancemax. 10 Ω for each connection
Line fault detectionlead monitoring for all lines and CJC
Transfer characteristics
Resolution16 Bit
Step response65 ms (0 ... 90 % of the measured value by smallest filter setting)
Directive conformity
Electromagnetic compatibility
Directive 2004/108/ECEN 61326-1:2006
Explosion protection
Directive 94/9/ECEN 60079-0:2006, EN 60079-11:2007, EN 60079-26:2004 , EN 61241-0:2006, EN 61241-1:2006
Standard conformity
Insulation coordinationEN 50178
Electrical isolationEN 60079-11:2007
Electromagnetic compatibilityNE 21:2006
Protection degreeIEC 60529
Climatic conditionsIEC 60721
Ambient conditions
Classification3K3
Ambient temperature-20 ... 70 В°C (-4 ... 158 В°F)
Storage temperature-20 ... 100 В°C (-4 ... 212 В°F)
Relative humidity95 % non-condensing
Shock resistance15 g peak, 11 ms period
Vibration resistance2 g , 10 ... 500 Hz according to IEC 60068-2-6
Damaging gasacc. to ISA-S71.04-1985, severity level G3
Mechanical specifications
Connection typeTerminals
Core cross-sectionmax. 2.5 mm2
Protection degreeIP20, for on-site installation a separate housing is required with a minimum of IP54
Massapprox. 245 g
MountingDIN rail mounting
Data for application in connection with Ex-areas
EC-Type Examination CertificateDMT 98 ATEX E 015 X , for additional certificates see www.pepperl-fuchs.com
Group, category, type of protectionEx-Hexagon II (1)2G EEx ia/ib IIB/IIC
II (1D)(2D)
Temperature classT4
Supplyonly in connection with the power units RSD2-PSD2-Ex4.34, RSA6-PSD-Ex4.34
Input
Voltage9 V
Current38 mA
Power86 mW
External capacitance4.9 µF
External inductance20 mH
L/R-ratio0.4 mH/Ω
Internal buscustomer specific
Statement of conformity
Group, category, type of protection, temperature classEx-Hexagon II 3D IP54 T 90В°C
Electrical isolation
Input/inputno electrical isolation
Input/power supplysafe electrical isolation acc. to EN 60079-11:2007, voltage peak value 60 V
Input/Internal Bussafe electrical isolation acc. to EN 60079-11:2007, voltage peak value 60 V
Internal bus/power supplysafe electrical isolation acc. to EN 60079-11:2007, voltage peak value 60 V


Назад в раздел